Wireless Design Considerations for Medical Devices

INTRODUCTION

Wireless connectivity for medical devices is no longer a long-shot dream, but an expectation from patients. Wireless devices are complex; however, with some planning, major risks can be avoided. Below we outline an introduction to the general categories and delve into specific examples. The broad categories include:

  • Intended Application
  • Intended Region of Operation
  • Medical Regulations
  • Wireless Regulations & Certifications
  • Miscellaneous Regulations
  • Additional Design Considerations
  • Manufacturing Considerations

These categories should be viewed together, since they can influence each other. We can start by asking specific questions to better understand the solution process.

INTENDED APPLICATION

The intended application includes the specific problem, proposed solution, and target population.

What is the specific problem?

What is the proposed solution?

Who is the target population?

The previous three questions help shed light on what wireless design approach is viable. Asking more questions may help increase awareness. For example, what is the technical aptitude of the target population? Must all wireless components work autonomously without user interaction? How should the user interact with the system?

INTENDED REGION OF OPERATION

What countries is the device expected to operate?

Of course, the operating country will impact the appropriate medical and wireless regulations. For example, the FDA is the regulatory body of medical devices marketed in the United States. Also, because of FCC regulations, the devices’ operating frequency ranges will be limited and, therefore, a design factor. For example, in the case of LTE CAT 1, frequency Band 2 (downlink 1960 MHz and uplink 1880 MHz) can be used in the US [1]. However, in Europe, Band 2 is not permitted, and Band 28 must be used (downlink 780.5 MHz and uplink 725.5 MHz). If the device will be operated in both regions, one could either choose bands that are common to the target locations (ex. Band 1 – downlink 2140 MHz and uplink 1950 MHz) or provide a separate software configuration that is chosen depending on the location.  The frequencies can impact the size requirements of the circuit, since lower frequencies with multiple bands tend to take more space than a single banded high frequency circuit. So, tradeoffs between the operating bands/frequency and size (as well as operating power) are ubiquitous.

What about operating device user proximity?

Specific health guidelines also include safe distance of the wireless device from human tissue. For example, the Specific Absorption Rate (SAR) measures the rate at which RF energy is absorbed by the body. SAR testing uses models of the human body that are filled with liquids to simulate human tissue RF absorption [2]. For the frequency bands of interest, the SAR values are tested at the most severe (not necessarily typical) operating conditions. Therefore, in some cases, the SAR value may pertain to a position or direction that is seldomly used.

MEDICAL REGULATIONS

Considering the medical regulations as a whole package helps ensure nothing is missed in the early product development phases. Some regulations include:

StandardTitle
ISO 13485Medical devices – Quality management systems
ISO 14971Application of risk management to medical devices
IEC 60601Medical electrical equipment requirements
IEC 62304Medical device software – software life cycle processes

Ensuring that adequate medical device quality standard processes (ISO 13485) are in place prior to development is key. In terms of risk management (ISO 14971), identifying, documenting, and mitigating risks is paramount. So, addressing any Specific Absorbance Rate (SAR) concerns for wireless products is an example risk captured in the risk management process. Additionally, understanding how IEC 60601 requirements would impact the wireless design is essential. For example, if an audible alarm is required for IEC 60601 to warm of an imminent failure, must that capture any failures associated with the wireless components?  Finally, understanding how the software aspects of medical device certification in the context of the hardware component selection is important. What specific wireless functions must be certified to what specific software class level? The answer clearly depends on the consequence of the component’s failure.

As a result, all these aspects are interconnected and should be analyzed together.

WIRELESS REGULATIONS & CERTIFICATIONS

What regulatory body(s) is/are required?

The FCC is the regulatory body in the United States that specifies whether a specific device can operate at a specific frequency with a specified power level in specific directions for a designated application.

  • FCC Part 15B: Unintentional Radiator
  • FCC Part 15C-F, H: Intentional Radiator

The unintentional radiator categoryspecifies the acceptable power levels for frequencies operating between 9 KHz and 3 THz but not “intended to emit RF energy wirelessly.” [3]

This applies to, for example, an onboard microcontroller that has a CPU operating at 1 MHz. In this case, even though wireless power is not intentionally transmitted, there is still generated RF energy (due to Maxwell’s equations). The acceptable power levels in this category are generally lower than the intentional radiator part and, therefore, still require electromagnetic interference (EMI) minimization techniques.

The intentional radiator category specifies the acceptable power levels for frequencies intended to be emitted. Full certification to this category can be mitigated if using an antenna and network like one already found to be compliant; this observation, therefore, can reduce development costs.

For both categories, specific design considerations should be used to mitigate these risks. This is addressed typically in the mechanical packaging and, most importantly, the Printed Circuit Board (PCB) layout portions of the design process.

What are some additional certifications?

Additional certifications will depend on the specific use case. For example, in the case of cellular Internet of Things (IoT), the 3rd Generation Partnership Project (3GPP) standards for cellular specifies that the allowed circuit voltage must be at least a specified voltage. Also, adherence to the PCS Type Certification Review Board (PTCRB) certification may be required by various cellular carries, like Verizon. This observation is critical, since in connected care applications, maintaining compliance to suppliers’ requirements may be overlooked but result in integration risk. Of course, other certifications may be applicable depending on the application.

MISCELLANEOUS REGULATIONS

Depending on the specific application, other regulations may apply. For example, the Health Insurance Portability and Accountability Act (HIPPA) in the United States protects sensitive patient health information. Also, the General Data Protection Regulations (GDPR) are a set of compliance regulations that protects citizens of the European Union.

WIRELESS DESIGN CONSIDERATIONS

What is the approach for wireless antenna?

Choosing the topology of the antenna is not trivial and a critical system design choice; a full treatment of the subject is out of scope. Instead, several guiding principles will be mentioned. In the case of a 2.4 GHz application, most antennas follow three general approaches – 1) Wire Antenna, 2) PCB Antenna, and 3) Chip Antenna.

Antenna TypeSizeCostEfficiencyEase of Manufacturing
WireGreatestGreatestGreatestLowest
PCBMiddleLowestLowestGreatest
ChipLowestMiddleMiddleMiddle

So, if size is a constraint, a chip antenna may be best. If ease of manufacturing is important but not efficiency, a PCB antenna could be suitable. If efficiency must be optimized over all other variables, a wire antenna can be a viable option.

What is the approach for wireless antenna tuning?

Consider the following topics:

  • Ground Clearance around antenna
  • Optimal Antenna Placement
  • Antenna Feed Consideration
  • Antenna Matching network

In terms of the antenna feed consideration and antenna match network, maximizing the power delivered to the antenna by minimizing reflections is a commonly employed technique in wireless design. A common tool used in the technique by RF engineers is the Smith Chart, as shown below. Fundamentally, the impedance is measured at the frequency range of interest, plotted on the chart, and modified by using capacitors, inductors, and, in some cases, resistors. The goal is to move the impedance to the middle of the diagram (labeled “Matched Impedance”).  

Fig. 1: Simplified Smith Chart

The process of tuning is to ensure the impedance from the perspective of the integrated circuit (IC) is equal to the impedance from the perspective of the antenna and equal to the characteristic impedance of the RF trace. Otherwise, significant reflections will result in power dissipation, and therefore, significantly reduce the distance of wireless operation.

Return Loss (dB)Power Reflected %Power Delivered to Antenna %
0.0199.770.23
0.197.722.28
179.4320.57
101090
20199

As the previous table demonstrates, due to conservation of energy, the more energy that is reflected, the less useful power is delivered into the antenna, degrading the performance of the overall system. Therefore, tuning the antenna is a key element of the wireless design process.

Specifically, the following points can simplify the tuning process:

  1. Calibrate network analyzers prior to tuning.
  2. Use only high-Q components.
  3. Ensure capacitors have a series resonance at least double the operating frequency.
  4. Ensure inductors have a self-resonance at least double the operating frequency.
  5. Shunt components should be on the RF trace.
  6. Measure impedance at the same location at which components will ultimately lie.
  7. If multiple bands will be operated, tune the lower frequency band first.

Also, to help minimize the EMI emissions and simplify the PCB layout process, the following PCB layout is recommended [4]:

Fig. 2: Four Layer PCB Stackup

On the other hand, two-layer PCBs may be used in some cost-constrained applications but make the PCB routing more difficult because their characteristic impedance is directly proportional to the substrate height and would, therefore, require thicker RF traces. For completeness, an example 2-layer stackup could be considered:

Fig. 3: Two Layer PCB Stackup

ADDITIONAL DESIGN CONSIDERATIONS

What other non-wireless functions are required?

Considering the wireless function requirements in the context of the non-wireless requirements is important.

Fig. 4: Wireless & non-wireless function separation

The wireless requirements may dictate that a specific chipset with certain set of characteristics be used. But, only a subset of those specific chipsets may address the non-wireless functions as well. In the case of a medical device, a portion of the system will be used to perform some diagnostic or treatment operation. This may be performed with non-wireless components. The transmission of the data of interest (ex. breathing rate, medication status, etc.) will be performed by a wireless function component. The delineation as well as interactions between these two subsystems is a critical design choice – what is the best interface?   

What are the space requirements?

In the ideal case, the mechanical packaging is designed around the antenna, not the other way. Otherwise, compromises on the size may negatively impact system performance. Therefore, the space requirements highly impact the system design. In practice, however, there are physical constraints. The specific dimension that is constrained impacts the type of antenna that may be used.

What are the mechanical provisions?

The mechanical casing can impact the effective dielectric constant of the transmission media, from the perspective of the antenna. Ergo, understanding the mechanical casing and placing it nearby the antenna during tuning is strongly recommended.

MANUFACTURING CONSIDERATIONS

For manufacturing, there are various common categories to consider; three categories include design-for-manufacturing (DFM), part obsolescence, and manufacturing quality.

Design-for-Manufacturing considers the set of criteria to minimize product failures by maximizing the quality of the design decision process. This may include:

  • Tenting PCB vias wherever possible
  • Allowing an additional space clearance between PCB elements beyond the minimum mandated by the board manufacturer
    • An example is keeping traces as far apart as practical and keeping traces away from the edge of the board (to limit board edge oxidation).

Of course, there additional criteria beyond these examples.

Parts obsolescence is another consideration. Wherever possible, choosing electronic components with common PCB footprints and electrical properties in the event of part obsolescence reduces significant change control process rework. General part obsolescence risk mitigation can be documented as part of the company’s ISO 14971 risk management process for the project. 

Manufacturing quality can be decomposed into PCB board manufacturing and PCB assembly. Both components of the process are critical to ensuring adequate quality control.

In terms of PCB Assembly, fabrication to the IPC-A-610D Class 3 is also recommended for safety critical applications including medical devices.

Certifying to an IPC class 2 standards allow for extended life when compared to Class 1 but does not ensure uninterrupted service. If continuous operation of the wireless portion of the system is not as critical in the application, Class 2 may be possible. Furthermore, IPC-A-600 covers the PCB board manufacturing, itself. Note that there are additional standards in PCB manufacturing.

Fig. 5: IPC-A-600 & IPC-A-610 Simplified Relationship

Additionally, the manufacturing quality process must be documented consistently and integrated with the medical product quality management system, like the flow of Fig. 6.

Fig. 6: Manufacturing Quality Process

REFERENCES

[1] – Haltian Global IoT Frequency Bands E-Book

[2] – Specific Absorption Rates for Cell Phones – https://www.fcc.gov/sites/default/files/sar_for_cell_phones_-_what_it_means_for_you.pdf

[3] – FCC parts – https://www.fcc.gov/oet/ea/rfdevice

[4] – Cypress Antenna Design and RF Layout Guidelines

Tutorial: Medical Device Design with MATLAB Simulink MBSE

We will discuss an emerging system design technique that has been proven to reduce complex product development costs by 55% [1].

INTRODUCTION & WHY?

As product complexity in virtually every industry has increased exponentially, processes to design and prove device correctness have become more involved. Traditional methods for complex system design in highly regulated industries, such as medical device and aerospace products include the following (simplified) steps1:

1). Manual gathering of requirements
2). Design system and/or architecture
3). Coding and/or development
4). Validation & Verification

Typically the process outlined above is performed in what is referred to as a waterfall approach, in which each of these steps is followed sequentially and fully. Particularly, airborne software standards follow the DO-178C standard and medical software components follow the life cycle processes outlined in IEC 62304.

Suppose during step (4), (validation), a missing requirement is caught. Adding the requirement would require redesign, re-coding, re-verification, and re-validation. This is one reason highly regulated products oftentimes require greater capital investments than consumer electronics.

One seemingly obvious solution is to ensure requirements are fully correct and comprehensive in the early product development stages. In practice, achieving such a feat isn’t always realistic (especially if the product is complex) due to a gamut of reasons including limited foresight, evolving interfaces controlled by external groups, and dynamic regulatory requirements (think General Data Protection Regulation (GDPR)).

Another solution (albeit a poor choice) could be to attempt abridging certain subsets of the process. However, such an endeavor would prove to be haphazard, error-prone, and likely to reduce product quality. An alternative to the above quandary is not necessarily attempting to avoid the process but rather shortening the burden of re-running the steps through automation; Model-Based System Engineering (MBSE) is one such approach.

WHAT IS MODEL BASED SYSTEM ENGINEERING (MBE)?

MBSE is a system design technique that decomposes a system as a representation of simpler elements (via models), connects the models together, and continues to represent complex models in terms of smaller ones. The following terms make up a “system,” as [2] excellently defines:

  • Entity are components that make up a system
  • Attributes are characteristics of entities, like temperature or volume
  • Relationships are associations between attributes and entities based on causality

The key observation is that a system can be broken down hierarchically. So starting with the highest level of abstraction, the top level (system) is designed first and broken into its elements, namely subsystems. The process is repeatedly performed until the lowest level is sufficiently simple and therefore requires no additional decomposition. The definitions of these levels (in progressively greater granularity) are as follow:

  • System
  • Sub-systems
  • Assemblies
  • Sub-assemblies
  • Parts
Fig.1: System Decomposition

HOW TO PUT MBSE INTO PRACTICE?

We can follow a process similar to the formal System Analysis and Design Technique (SADT). Classically, engineers and scientists use these methods by drawing and breaking down the system on a whiteboard or a piece of paper. However, with advances in computer-aided (CAD) modeling, system decomposition can be performed on a computer, using tools, such as CAMEO or MATLAB Simulink. After the design phase is completed, the model is verified, and the code is automatically generated. So, the burden of writing software is significantly reduced.

Consider the impact in time savings. Instead of having to create a 1 million-line program, the vast majority of the coding could be automated. The savings is further underscored when feature enhancements are considered in the development cycle (typically via a change control process). Therefore, effort impacts from required changes are greatly reduced. Faster product cycles (weeks instead of years) becomes possible.

REQUIREMENTS ANALYSIS & GATHERING

For the purposes of this tutorial, we’ll use MATLAB Simulink to design part of a mechanical ventilator medical device. When we start with a medical product, generally we gather product requirements, which include regulatory guidelines. IEC 60601 generally provides additional guidance, such as a required medical device’s alarm volume in decibels and expected operating modes. For this exercise, we will specify a simplified set of requirements, as shown in Fig. 2.

Fig. 2: Requirements Subset

With Simulink modeling, the requirements in Fig. 2 can be linked to the model itself, with the following steps accompanying Fig. 3:

  1. Open the requirements document in word and highlight the requirements text
  2. Right-click a specific Simulink block
  3. Select Requirements
  4. Select Link to Selection in Word
Fig. 3: Linking Requirements to Model

Fig. 4 shows how to ensure the requirement mapping is successful.

Fig. 4: Linked Requirements to Model

DESIGN INTERFACES & HIGH LEVEL

Let us breaking our system in three main components:

  1. User Input (Pre)Processor
  2. Tube & Patient Lung
  3. Ventilator
    • Alarm Subsystem
    • Control Micro-controller Unit (MCU) Subsystem
    • Pneumatic Subsystem
      • Sensors
    • Pneumatic Controller

The Ventilator will take user input from the (pre)processor/panel and accordingly adjust the output airflow (for example) to the patient. Inside the ventilator, there is a primary control MCU that controls the Pneumatic Controller, which drives the Pneumatic subsystems (which, in turn, includes sensors and pneumatic components). The alarm subsystem will make a noise and visual indicator if certain fault conditions, such as power loss, are detected.

Consistent with the above, the Simulink Model Browser shows the following:

Fig.5: System Decomposition

At the highest level, we design the following interface and major entities comprising the system.

Fig.6: High-level Block Diagram & User Interface Simulation

DESIGN LOW LEVELS

Next, we will dig into the yellow box of Fig. 6 and create a Control Logic MCU, Alarm Subsystem, Pneumatic Subsystem, and Pneumatic Controller.

Fig. 7: 2nd Level Decomposition

We continually decompose the system until the constituent parts are sufficiently simple and thus no longer divisible. Additionally, the primary control logic implementing the requirements of Fig. 2 is shown in Fig. 8.

Fig. 8: 3rd Level Decomposition – Control Sequence of MCU

For tutorial succinctness, we’ll stop at the third level.

SIMULATION RESULTS

Of course, testing the model is paramount. Testing phases before and after automatic code generation are commonly employed. For now, we will test the model itself. As shown in Fig. 9, the system was able to detect fault conditions when the ventilator was turned off abruptly during ventilation mode (shown visually with the red alarm indicator when the system ON input turned red as well).

Fig. 9: User Turns Machine off during Patient Ventilator – Fault Detected

By running the simulation while the state machine window is open, the control logic’s state machine can be debugged using Simulink’s Stateflow. The Blue boxes represent the current state machine’s state.

Fig. 10: Video Debugging & simulation of state machine

For specific plots pertaining to the patient’s flow, pressure, and volume, Fig. 11 shows the simulated readings from the ventilator’s sensors.

Fig. 11: Image of MBSE-designed Ventilator Patient Simulation

And, we conclude testing with a video showing the readings in live-action.

Fig. 12: Video of MBSE-designed Ventilator Patient Simulation

AUTOMATIC C/C++ CODE GENERATION

As mentioned, a prodigious added benefit of MBSE is automatic code generation, which has saved significant product development costs by ~55% [1]. Code generation is an excellent topic for a future blog article.

NEXT STEPS & ENHANCEMENTS

Verification (and coverage) should be performed as well. An example test includes mathematically integrating the output flow every cycle to ensure the tidal volume is truly delivered to the patient within an error margin. An assertion can be tied to this specific condition for every breathing cycle (often referred in computer science as an invariant).

FOOTER COMMENTARY

1 Implementing a commercial product generally requires several additional steps, such as design reviews. In the context of medical products, risk management processes and quality management processes, for example, need to be in place before, during, and after the product development process. ISO 14971 and ISO 13485 standards, although not directly discussed in this article, would, therefore, be required.

2 Depending on the product’s safety classification (design assurance), the MBSE elements (and the design tool itself) would need to be qualified, which means proven to be developed with a sufficiently high confidence level for meeting the product’s requirements.

REFERENCES

[1] MBE Cost Benefits (PTC)

[2] MBSE Primer

ABOUT SIMPLONICS

Simplonics is a leading electronics design consulting and a supplier firm for companies across the United States in highly regulated industries, including medical devices. Our common services include electrical hardware design, software development, and sensor system design.

Our customers tell us horror stories about ex-vendors who delivered defective systems, resulting in safety concerns, vulnerable software, and other issues. We’ve successfully helped companies out of these situations by consistently delivering superior quality and simultaneously reducing recurring costs by more than 20%.

Are you working on a technical challenge and want a fresh perspective? Feel free to reach out by visiting the Contact Us page and submitting your contact information. Let’s build a great relationship for the future.

Tutorial: How to Detect Cancer Cells with Machine Learning?

The medical device industry has been advancing towards solving diagnostic and treatment problems with machine learning (ML), which is a data prediction technique. Therefore, investing an effort to understand this multidisciplinary area for your own application can help you maintain an edge over your competition. To help you accomplish that end, we’ll cover a relevant case study for (ML) by first defining necessary terms, going into some theory, and implementing a Python coding example.

Conventional methods of data predictions use statistical techniques, such as regression, to classify data points or predict future values. The increase of computers’ computational power has made solving sophisticated algorithms with intense operations possible. In this tutorial, we will study and implement a Support Vector Machine (SVM) technique to categorize whether medical tumor cells are cancerous by studying their features; the principles of this tutorial can be applied to ubiquitous classification problems.

DEFINITIONS

Let’s start with a few definitions. Two common types of machine learning technique are supervised learning and unsupervised learning.

Supervised Learning– When the correct categories pertaining to input data points are known. The Support Vector Machine (SVM) is an example we’ll be studying here.

Unsupervised Learning – Occurs when the output targets aren’t known in the given problem. We would analyze commonalities among the data itself to find groupings of similar data together.

Features (Inputs) – Specific input that map to an output class target. A cell’s mean area and mean smoothness are two examples we’ll study here.

Target (Output) – “Correct” answers (determined classes) pertaining to the specific feature inputs.

Training Set – Subset of data used to build machine learning model. These data points are not used in the testing stage.

Test Set – Subset of data used to determine accuracy of model. These data points are not used in the training stage.

Class – Categories to which the input features pertain.  In this example, Malignant and Benign are the two possible classes for tumor cells. Other applications may have more than two possible classes.

Interference – Using the trained ML model, deduce to which class a test input pertains.

Margin – Distance between closest points of different classes in the context of Support Vector Machine. The support vectors are simply the points closest to the opposing class. During training, the support vectors are computed to determine the hyper-plane (in sufficiently high dimensions). Fortunately, after training, almost all data points can be disposed and only the support vectors are retained, resulting in significant storage space reductions.

THEORY

Support Vector Machines (SVMs) are a type of supervised learning algorithm that attempts to find a dividing line/curve (or hyper-plane in higher dimensions) so that unknown data points can be categorized in the appropriate class. It’s best to illustrate with some diagrams

Fig.1: Data points with Features 1 and 2 Plotted

Clearly, a line between the data points for the two classes (X’s and O’s) would serve as a reasonable divider for the data points. But, what’s the equation of that line? And what does it look like in higher dimensions?

Fig.2: Data points with Features 1 and 2 Plotted after SVM Invocation

The goal is to find where to draw the thick red line above in Fig. 2. Our goal is to maximize the margin. The data points (X’s and O’s above) closest to the thin red lines are called the support vectors.

The example above appears relatively simple and may not require using the SVM technique. So, why is SVM useful? It becomes useful when the data points don’t appear to be linearly separable, which means separation with a single decision surface. Because we are effectively solving for an equation that separates the data, transforming a low dimensional non-linearly separable to a higher linearly separable one will simplify the solution. The function used for this purpose is a kernel function, which is used to transform input data to higher dimensions.

1). Linear (‘linear’)

2). Polynomial (‘poly’)

3). Radial (‘rdf’)

Mathematically, we can write the SVM training equation, according to [1]:

In Eq [1] above, K is the kernel function, x is a matrix containing inputs we’d like to train, t represents targets, and the second term is added to help make the equation linearly separable in higher dimensions. We’ll use the Sklearn [2] library in python solve this equation for us. Other packages, such as cvxopt [3], would use a form similar to Eq [1], whose form is the same as the Lagrange Multiplier solutions.

IMPLEMENTATION

1. Import Libraries

First, we import the sklearn, numpy, matplotlib, and math libraries into our Python program.

from sklearn import svm
import numpy as np
import matplotlib.pyplot as plt
from sklearn.datasets import load_breast_cancer
import math 

2. Load Data

Secondly, we’ll load the breast cancer data set and also calculate the number of data points we have. We’ve got around 569 samples.

	dataset = load_breast_cancer()   
    sampleSize = dataset.data.shape[0] #sample size
    trainSize = math.floor(0.9*sampleSize) #90% of dataset is used for training
                                           #Thus, remaining 10% used for testing

3. Select Featuers

Next, we need to select a couple features to analyze.

#Choose fourth and fift columns as features 1 and 2, respectively
    #Off by one because of zero indexing
    feat1Index = 3
    feat2Index = feat1Index + 1
    feat1Name = (dataset['feature_names'][feat1Index])
    feat2Name = (dataset['feature_names'][feat2Index])

4. Structure Data for SVM Input

Additionally, we’ll have three sets of variables housing our data to make the example clear. First, we’ll get all of the data, then we’ll designate about 90% of our data for training, and the rest will be reserved for testing. For analysis and plotting purposes later, we further split the data depending on whether the target is malignant or benign (XMal and XBen, respectively).

    [f1, f2, y] = sliceData(dataset, 0, sampleSize, feat1Index, feat2Index) 
    #all data 
    X, XBen, XMal = separateFeaturesViaClasses(f1,f2,y)
    
    
    [f1Tr, f2Tr, yTr] = sliceData(dataset, 0, trainSize, feat1Index, feat2Index) 
    #train data 
    XTr, XBenTr, XMalTr = separateFeaturesViaClasses(f1Tr,f2Tr,yTr)
    
    [f1Te, f2Te, yTe] = sliceData(dataset, trainSize, sampleSize, feat1Index, 
                                  feat2Index) #Test Data
    XTe, XBenTe, XMalTe = separateFeaturesViaClasses(f1Te,f2Te,yTe)
def separateFeaturesViaClasses(f1, f2, y):
    # Creates and returns TWO (2) separate input features matrices - each 
    # pertaining to one of either target classes as well 
    # ONE (1) input features matrix pertaining to both target classes
    assert((len(f1) == len(f2) == len(y)))
    #Create scatter plot inputs for each class
    X = [[f1[i],f2[i]] for i in range(len(f1))]
    XBen = np.array([X[i] for i in range(len(f1)) if y[i] == 1]) 
    #Class 1 - Benign  
    
    XMal = np.array([X[i] for i in range(len(f1)) if y[i] == 0])
    #Class 2 - Malignant
    
    return X, XBen, XMal
def sliceData(dataset, start, end, feat1Index, feat2Index):
    #Slices features and output arrays based on indicies
    f1 = dataset.data[start:end,feat1Index]
    f2 = dataset.data[start:end,feat2Index]
    y = dataset.target[start:end] #same as the outcome ("Correct Answers")
    return f1, f2, y

5. Invoke SVM Algorithm

To have Python solve Eq. [1] for us, we’ll need to provide our training data set and correct target labels.

    #Fit the input parameters to an SVM model. Assume a linear kernel
    #We only want to provide the training data so we'll have some 
    #left for testing
    clf=svm.SVC(kernel='linear')
    clf.fit(XTr,yTr)

6. Analyze Results

We’ll set the accuracy to the ratio of correct test outputs divided by the total number of test attempts. We’ll see that we got 2 samples wrong out of about 60 test attempts.

    #Now we perform the inferencing step and analyze accuracy results
    modelOutput = clf.predict(XTe)
    correctOutput = y[trainSize:]
    result = modelOutput == correctOutput
    #get indices for misclassified samples
    wrongIndices = [i for i in range(len(result)) if (result[i] == False)] 
    xWrong = np.array(XTe)[wrongIndices]
    accuracy = sum(result)/len(result)
    accuracyStr = "Accuracy is: " + str(round(accuracy*100,2)) + "%"
    print(accuracyStr)

7. Plot Data

Lastly, we plot our data. We also draw the SVM decision curve by extracting the line’s slope and intercept points.

  # Calculate SVM Curve for plotting 
    w = clf.coef_[0]
    a = -w[0]/w[1]
    xx = np.linspace(650,700)
    TERM = (clf._intercept_[0]/w[1])
    yy = a*xx + TERM
    plt.plot(xx,yy)
    
	#Plot Data points
    plt.scatter(XBenTr[:,0],XBenTr[:,1], label='Benign - Train Data',
                marker='o', color='blue')
    plt.scatter(XBenTe[:,0],XBenTe[:,1], label='Benign - Test Data', 
                marker='o', color='orange')
    plt.scatter(XMalTr[:,0],XMalTr[:,1], label='Malignant - Train Data', 
                marker='x', color='blue')
    plt.scatter(XMalTe[:,0],XMalTe[:,1], label='Malignant - Test Data',
                marker='x', color='orange')
    plt.scatter(xWrong[:,0],xWrong[:,1], label='Incorrect Test Outputs', 
                marker='+',color='red')
    plt.legend()
    plt.xlabel(feat1Name)
    plt.ylabel(feat2Name)
    plt.title("Support Vector Machine Example for Cancer Cell Classification")
    plt.text(400, 0.22, accuracyStr,bbox=dict(facecolor='red', alpha=0.5))
    plt.show()

Below, we have the plot from our work. We achieved about a 96.5% accuracy.

Fig.3: SVM Model Performance – 96.5% Accuracy

NEXT QUESTIONS

In production, we would optimize our accuracy further and consider the computation resources for the training and inference stages. Here are some questions to consider.

  1. How does varying the kernel function affect performance?
  2. How would the code example be modified to accommodate higher dimensions, such as three features? Would that change improve accuracy?
  3. What features are optimal for the above problem?
  4. How does the training and inference time grow with the number of features? Does this agree with theoretical estimates?
  5. What’s the optimal value of gamma and C, as defined in [2]?

REFERENCES

[1] – Machine Learning An Algorithmic Perspective. 2nd Edition. Stephen Marsland.

[2] – https://scikit-learn.org/stable/modules/generated/sklearn.svm.SVC.html

[3] – https://cvxopt.org/

Let us know when you’d like to discuss how the learning in this tutorial may be applicable to the technical problem you’re trying to solve. Our fresh view of your problem may give you a different, valuable perspective to consider.

How to Design Medical Devices with Dynamic Regulations?

With advances in the technology industry, instantaneous data access is not only ubiquitous, but also expected in the medical device industry. Health delivery and monitoring services are no longer solely in healthcare facilities; these services are becoming more pervasive in the patients’ homes, shifting the medical treatment paradigm and creating opportunities for new disruptive electronic technologies. The shift also engenders challenges to prevent compromise or misuse of patient data. In this blog, we’ll show how one can begin designing a medical device in the context of certain regulations to minimize rework between prototype and production design builds.

Consideration #1: What is a process for designing medical device electronics?

System engineering, which consists of understanding and defining the system requirements (including regulatory and customer expectations) are paramount to the design process. Federal regulations for both medical devices (and even aerospace systems) both regard traceability as essential. Traceability is a relationship mapping between two or more products of the development process. Increased traceability rigor is required as the criticality of the medial device is increased (Class III vs Class I, for example).

We traditionally design systems from the top-down, which means that we look at the overall system perspective and consider all the available requirements from all stakeholders. So, requirements are allocated to appropriate software or hardware designations, broken down into increased specificity, and implemented. Validation means Do you have the right requirements? while verification asks the question does your implementation meet the requirement? This process is called V&V (Validation & Verification).  

Requirements Validation 
Requirements Verification 
Time

Design and verification occur in opposite directions:

3 
2 
Design (Top Down) 
1 
1 
2 
Verification (Bottom Up) 
3 
Lowest Level Subsystem Module Highest Level Subsystem Module 
Order in Design Process 
Lowest Level Subsystem Module Highest L I Subsystem Module 
Order in Verification Process

Design is performed from the top-down, which means that high level software modules are defined before lower level ones. Lower level modules are combined to make up one or more higher level modules. Lower-level modules are verified first to mitigate integration risks. It’s recommended to perform requirements-based testing before system-level testing.

Consideration #2: What are some landmines to avoid when designing new medical devices?

Functional requirements, for example those pertaining to customer expectations, as well as regulatory requirements, must be considered simultaneously in the design process. Even the initial goal is to develop a prototype demonstrating the principle for your stakeholders, neglecting to consider the holistic system perspective by asking who are all of my stakeholders? And what do they care about? can lead to rework. This is because hardware or software components chosen to meet the functional requirements may not meet regulatory ones. So, for example, embedded computers (and other components) may have to be replaced in the production design. How much additional effort would we inadvertently append to the overall hardware and software development effort? 20%? 40%? From our experience, the cost delta can range from 30% – 50%.

Ephemeral goals without long-term design strategy context likely increases overall development cost. It’s a common observation we’ve seen with the myriad companies working in this space. Let’s study how certain regulatory requirements rework can be avoided for a medical device example by considering all critical elements sufficiently early in the design process.

Consideration #3: What are some example regulations to consider?

According to the Food and Drug Administration (FDA), regarding medical records, Title 21: Food & Drugs – Part 21 – Protection of Privacy (21.71 (5)) states that a “record is to be transferred in a form that is not individually identifiable.”  In another statement, (21.70 (a)) states that “Names and other identifying information are first deleted and under circumstances in which the recipient is unlikely to know the identify of the subject of the record.” [1]

To paraphrase, transmission and storage of customer records must be performed in a way that protects patient confidentiality. Simply transmitting the patient’s name and storing the name and data in one place won’t put federal regulators at ease. Suppose we’re designing a subsystem called the Diagnostic Reporting Function (DRF) of our next medical device, and we want to comply with the Part 21 regulations above. Perhaps we would enforce a requirement on our medical device, such as:

Diagnostic Reporting Function (DRF) shall generate a unique identifier for each heart-rate monitor reading. [REQ-1]

Diagnostic Reporting Function (DRF) shall transfer the generated unique identifier along with the recorded heart-rate monitor reading to the Transmission Endpoint System (TES). [REQ-2]

Medical Device 
Diagnostic 
Reporting 
Function 
Internet 
Transmission Endpoint System 
Unique 
ID/Patient 
Isolator 
Customer/Patient

We won’t get into too much detail in the above figure, but it’s a basically a underscores how one may decouple identifiable patient information with patient medical data to make it harder for unauthorized individuals to view sensitive information. Note that transferring patient data within an organization may be acceptable, if the Transmission Endpoint System (TES) is operated by a single organization and/or appropriate vendor risks are mitigated [1].

We have now decoupled health records from patient identification, a key regulatory requirement. Clearly this requirement impacts the software development of our example medical device. We also have other regulatory requirements to mitigate other cyber-risks. What about encryption? Sending the output of the medical device in plain-text (not encrypted) technically satisfies our above REQ-1 and REQ-2, but do not satisfy other regulatory requirements or guidelines. So, in the spirit of the V&V figure described previously, we have not yet satisfied our requirements validation process, since we need more requirement coverage.

Consideration #4:  What questions should we answer in order to develop a more secure medical device?

As an illustration, Remote Patient Monitoring (RPM) patients expect an intuitive user interface (UI) for their monitoring products. And, for device manufactures to deliver on this expectation safely, reviewing the overall system architecture is paramount to ensuring that security and privacy vulnerabilities risks are mitigated, according to a recent National Institute of Standards and Technology (NIST) article titled Securing Telehealth Remote Patient Monitoring Ecosystem [2].

Based on the NIST Cybersecurity Framework Version 1.1 [3], the same article ([2]) defines a set of Desired Security Characteristics guidelines, pertinent to medical device designers. Let’s pose the guidelines as questions to frame additional requirements for our medical device example (Note that these may be more relevant to the TES mentioned above):

  • Identify
    • How are network assets identified and managed?
    • What are the vendor risks, such as cloud providers or technology developers?
  • Protect
    • How are users identified and authorized appropriately with access control?
    • Is least privilege enforced for every user? 
    • Is data integrity and encryption (including data at rest) enforced?

The protect category especially echoes the principles outlined in the excellent book Zero Trust Networks: Building Secure Systems in Untrusted Networks by Evan Gilman and Doug Barth) [4].

  • Detect
    • What are the ways in which cyber threats are detected?
    • How are continuous security monitoring practiced?
    • In what ways are user account behaviors studied with analytics?
    • What type of information do security logs contain?
  • Respond
    • When a cyber threat event is discovered, what are the processes to limit the extent or limit of damage?
  • Recover
    • After a cyber incident, what are the processes to restore systems affected by the event?
    • How does the cyber incident recover process handle external and internal parties?

Consideration #5:  What is the impact of missing regulatory requirements in the design process?

Through the guidelines above, we can determine useful considerations to ensure systems are resilient against most common forms of cyber threats. Neglecting security best practices puts patient (and thus company reputation) in peril; in fact, there’s more than one cyber hack every 39 seconds, according to Michel Cukier of the University of Maryland [5]. To ensure the medical device we’re designing will be more prepared, let’s look at one question from the above – Is data integrity and encryption (including at rest) enforced?

Data integrity means that the data has not been tampered. A variety of common algorithms to check that the data has not been altered (mistakenly or intentionally) is checksum or hash. SHA-256 is an example.

So, what can some ancillary requirement be?

Some example, though not exhaustive, could be:

DRF shall transmit the computed SHA-256 of the message payload to TES. [REQ-3]

DRF shall encrypt all outbound communication to TES with the Transport Layer Security (TLS) version 1.2 or higher standard using the Advanced Encryption Standard (AES)-256 cipher. [REQ-4]

If we developed our medical device prototype without considering the full system view (ex: only REQs 1-2) of the end product, we may have (for example) chosen a microcontroller unit (MCU) that didn’t have the relevant encryption capability, resulting in prodigious hardware and software rework, and in our experience, may increase costs by more than 30-50%.

Fortunately, there is a way to mitigate these types of technical risks. And, we can help you accomplish both your short-term and long-term objectives through these disciplined processes. There are of course many more risks than we can ever write in this tutorial, and as you develop your product, we can help you avoid them. If you’d like us to brainstorm with you or if you’d like us to cover a future blog article topic, you’re welcome to comment or send us a message on our Contact Us page.

References:

[1] FDA Title 21: Food & Drugs – Part 21

[2] Securing Telehealth Remote Patient Monitoring Ecosystem       https://www.nccoe.nist.gov/sites/default/files/library/project-descriptions/hit-th-project-description-final.pdf

[3] NIST Cybersecurity Framework Version 1.1 https://nvlpubs.nist.gov/nistpubs/CSWP/NIST.CSWP.04162018.pdf

[4] Zero Trust Networks: Building Secure Systems in Untrusted Networks by Evan Gilman and Doug Barth https://www.oreilly.com/library/view/zero-trust-networks/9781491962183/

[5] Hackers Attack every 39 Seconds (NBC News) http://www.nbcnews.com/id/17034719/ns/technology_and_science-security/t/hackers-attack-every-seconds/